Sorry, you need to enable JavaScript to visit this website.
Skip to main content
Review
Classic

Systematic review of physiologic monitor alarm characteristics and pragmatic interventions to reduce alarm frequency.

Paine CW, Goel V, Ely E, et al. Systematic Review of Physiologic Monitor Alarm Characteristics and Pragmatic Interventions to Reduce Alarm Frequency. J Hosp Med. 2016;11(2):136-144. doi:10.1002/jhm.2520.

Save
Print
January 27, 2016
Paine CW, Goel V, Ely E, et al. J Hosp Med. 2016;11(2):136-144.
View more articles from the same authors.

Alarm safety is now a Joint Commission National Patient Safety Goal. This systematic review analyzed 24 studies on alarm characteristics and 8 studies that evaluated interventions to improve alert fatigue. Consistent with other studies, the vast majority of the time, alarms do not signal problems that require clinician action. The most promising intervention strategies for reducing alarms that have emerged thus far are widening alarm parameters, implementing alarm delays, and frequently changing telemetry electrodes and wires. A PSNet perspective discussed approaches to reduce alert fatigue while maintaining safety.

Save
Print
Cite
Citation

Paine CW, Goel V, Ely E, et al. Systematic Review of Physiologic Monitor Alarm Characteristics and Pragmatic Interventions to Reduce Alarm Frequency. J Hosp Med. 2016;11(2):136-144. doi:10.1002/jhm.2520.

Related Resources From the Same Author(s)
Related Resources