Sorry, you need to enable JavaScript to visit this website.
Skip to main content
Study

The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit.

Trbovich PL, Pinkney S, Cafazzo JA, et al. The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit. Qual Saf Health Care. 2010;19(5):430-4. doi:10.1136/qshc.2009.032839.

Save
Print
May 26, 2010
Trbovich PL, Pinkney S, Cafazzo JA, et al. Qual Saf Health Care. 2010;19(5):430-4.
View more articles from the same authors.

Errors at the administration stage are common for intravenous medications. Programmable or smart infusion pumps are widely used as a means of preventing such errors. However, prior studies have found that smart pumps alone may not significantly reduce errors, as they do not eliminate wrong-patient errors and may be prone to workarounds. This study compared three types of pumps—traditional pumps, smart pumps, and smart pumps combined with bar-code technology—in a simulated inpatient unit. The results indicate that smart pumps may reduce administration errors when combined with bar-coding or when only "hard" (unchangeable) dosing limits are used. Ultimately, creation of a "closed-loop" system that integrates technological solutions to prescription and administration errors represents the optimal solution for eliminating medication errors.

Save
Print
Cite
Citation

Trbovich PL, Pinkney S, Cafazzo JA, et al. The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit. Qual Saf Health Care. 2010;19(5):430-4. doi:10.1136/qshc.2009.032839.

Related Resources From the Same Author(s)
Related Resources