Sorry, you need to enable JavaScript to visit this website.
Skip to main content
Study

Novel analysis of clinically relevant diagnostic errors in point-of-care devices.

Shermock KM, Streiff MB, Pinto BL, et al. Novel analysis of clinically relevant diagnostic errors in point-of-care devices. J Thromb Haemost. 2011;9(9):1769-1775. doi:10.1111/j.1538-7836.2011.04439.x.

Save
Print
October 19, 2011
Shermock KM, Streiff MB, Pinto BL, et al. J Thromb Haemost. 2011;9(9):1769-1775.
View more articles from the same authors.

In this study, investigators compared international normalized ratio measurements (INR, a measurement of blood clotting ability) obtained simultaneously on a point-of-care analyzer and a standard blood draw. Although the concordance between the two measurements met traditional quality assurance standards, the point-of-care analyzer results were systematically biased toward normal measurements, putting patients at risk of preventable adverse events due to failure to adjust anticoagulant medications appropriately.

Save
Print
Cite
Citation

Shermock KM, Streiff MB, Pinto BL, et al. Novel analysis of clinically relevant diagnostic errors in point-of-care devices. J Thromb Haemost. 2011;9(9):1769-1775. doi:10.1111/j.1538-7836.2011.04439.x.

Related Resources From the Same Author(s)
Related Resources